FDA recall record

Getinge CM320 Series Washer Disinfector - Product Usage: A multi chamber washer disinfector recall — FDA Z-1124-2021

The potential that the plastic inlet hose connections may deform or break as a result of material fatigue and/or over tightening of the clamps. A broken…

Class IIclassification
Z-1124-2021official record ID
Feb 24, 2021report date

Recall record at a glance

Official recordZ-1124-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmGetinge Group Logistics America, LLC
Report dateFeb 24, 2021
Recall initiation dateDec 21, 2020
Quantity189 units total (Domestic 19 units, Foreign 170 units)
LocationWayne, NJ, United States

Official product description

Getinge CM320 Series Washer Disinfector - Product Usage: A multi chamber washer disinfector intended for washing, disinfecting and drying of surgical instruments, containers, operating theatre shoes and operating theatre accessories normally used in hospitals and central disinfection facilities. The disinfected items must be further processed through terminal sterilization before use in any surgical procedure.

Why the product was recalled

The potential that the plastic inlet hose connections may deform or break as a result of material fatigue and/or over tightening of the clamps. A broken hose connection may lead to a hazardous situation of water leakage or water spray. The water leakage or water spray may result in a burn to the user and/or an inconvenience due to a procedural delay if the Washer Disinfector is taken out of service.

Distribution information

Worldwide distribution - US Nationwide distribution including in the states of ND, PA, NC, VA, NE, MD, NY, MN and the countries of Australia, Austria, Belgium, Canada, Denmark, Finland, France, Germany, Hong Kong, Norway, Slovakia, Slovenia, South Korea, Sweden, Japan, China, Netherland, Italy, Kuwait and Czech Republic.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.