FDA recall record

Maquet Cardiovascular Us Sales Getinge 88-Series-washer disinfectors intended for cleaning recall — FDA Z-1475-2020

Potential leakage that may occur due to ruptured spot welds in the chamber, could generate a situation for the user such as risk for slipping or infection…

Class IIclassification
Z-1475-2020official record ID
Mar 18, 2020report date

Recall record at a glance

Official recordZ-1475-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMaquet Cardiovascular Us Sales, LLC
Report dateMar 18, 2020
Recall initiation dateFeb 10, 2020
Quantity146 US
LocationWayne, NJ, United States

Official product description

Getinge 88-Series-washer disinfectors intended for cleaning, disinfection and drying Model Number: 88-5 Part Number: 88-103; 88-203

Why the product was recalled

Potential leakage that may occur due to ruptured spot welds in the chamber, could generate a situation for the user such as risk for slipping or infection from contaminated water

Distribution information

Nationwide Foreign: Australia, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Estonia, Finland, France, Germany, Hong Kong, India, Iraq, Iran, Ireland, Israel, Italy, Japan, Kenya, Lithuania, Netherlands, New Zealand, Norway, Poland, Qatar, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.