FDA recall record

GENTIAN Cystatin C Calibrator and Calibrator Kit, for the following models: 1 recall — FDA Z-0099-2021

The firm has identified lots of calibrators that have shown to measure incorrect high values over time compared to the assigned values when released for…

Class IIclassification
Z-0099-2021official record ID
Oct 21, 2020report date

Recall record at a glance

Official recordZ-0099-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmGentian AS
Report dateOct 21, 2020
Recall initiation dateSep 2, 2020
Quantity107 calibrator kits
LocationMoss, N/A, Norway

Official product description

GENTIAN Cystatin C Calibrator and Calibrator Kit, for the following models: 1. Gentian C E, Cystatin C Calibrator, 1 x 1 mL, REF 1012, UDI: (01)07090038460092 2. Gentian C E Calibrator Kit, 6 x 1 mL, REF 1051, UDI: (01)07090038460108 3. Gentian C E, BCI Ref: A52763, Cystatin C calibrator Kit, 6 x 1 mL, REF 1051, (01)17090038460037 4. Gentian C E, Cystatin C Calibrator, 1 x 1 mL, REF 8012 (Not distributed in US)

Why the product was recalled

The firm has identified lots of calibrators that have shown to measure incorrect high values over time compared to the assigned values when released for distribution. This could result in higher Cystatin C values measured in patients.

Distribution information

US: MN, VA, KS,

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.