FDA recall record

Vero Biotech GENOSYL DS; Nitric Oxide Delivery System Combination Product NDA202860 recall — FDA Z-1863-2021

Reports received of NO measured below desired dose during transition between primary console and backup console since Software Version 2.2.3 was uploaded…

Class Iclassification
Z-1863-2021official record ID
Jun 30, 2021report date

Recall record at a glance

Official recordZ-1863-2021
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmVero Biotech, LLC
Report dateJun 30, 2021
Recall initiation dateMay 19, 2021
Quantity465 units
LocationAtlanta, GA, United States

Official product description

GENOSYL DS; Nitric Oxide Delivery System Combination Product NDA202860

Why the product was recalled

Reports received of NO measured below desired dose during transition between primary console and backup console since Software Version 2.2.3 was uploaded to consoles in the field.

Distribution information

Distributed nationwide to TX, MD, VA, GA, FL, KS, LA, MN, and NC

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.