FDA recall record

GE Medical Systems GE HealthCare OEC 9900 Elite C-arm Systems recall — FDA Z-2342-2025

GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted…

Class IIclassification
Z-2342-2025official record ID
Sep 3, 2025report date

Recall record at a glance

Official recordZ-2342-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmGE Medical Systems, LLC
Report dateSep 3, 2025
Recall initiation dateJul 2, 2025
LocationWaukesha, WI, United States

Official product description

GE HealthCare OEC 9900 Elite C-arm Systems.

Why the product was recalled

GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.

Distribution information

US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: FDA Mandated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records