FDA recall record

GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING GE Healthcare Infinia I Hawkeye 1 recall — FDA Z-0898-2025

GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear…

Class IIclassification
Z-0898-2025official record ID
Jan 22, 2025report date

Recall record at a glance

Official recordZ-0898-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmGE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Report dateJan 22, 2025
Recall initiation dateDec 18, 2024
Quantity93 units
LocationTirat Carmel, Israel

Official product description

GE Healthcare Infinia I Hawkeye 1, Model/Catalog Numbers: 1) H3000WN; 2) H3000YM; 3) H3000YS; System, Tomography, Computed, Emission

Why the product was recalled

GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Distribution information

Worldwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.