FDA recall record

GE Healthcare recall — FDA Z-2131-2016

GE Healthcare has recently become aware of a potential safety issue with the patient bore heating on the Discovery MR750w. A small area on the bore…

Class IIclassification
Z-2131-2016official record ID
Jul 13, 2016report date

Recall record at a glance

Official recordZ-2131-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmGE Healthcare, LLC
Report dateJul 13, 2016
Recall initiation dateJun 8, 2016
Quantity471 (184 US; 287 OUS)
LocationWaukesha, WI, United States

Official product description

GE Healthcare, Discovery MR750w

Why the product was recalled

GE Healthcare has recently become aware of a potential safety issue with the patient bore heating on the Discovery MR750w. A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used. No injury has been reported to GE as a result of this issue.

Distribution information

US: AK, AZ, CA, COM, CT , DC, FL, GA, GU, HI, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NV, NH, NJ, NY, NC, OH,OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WY. OUS: Algeria Argentina Australia Austria Bahrain Belarus Belgium Bermuda Brazil Bulgaria Canada China Egypt France Germany Hungary India Indonesia Ireland Italy Japan Kazakhstan Korea (Republic of) Kuwait Lebanon Mexico Netherlands New Zealand Poland Qatar Romania Russia Russian Federation Saudi Arabia Singapore Spain Sweden Switzerland Taiwan Thailand Turkey United Arab Emirates United Kingdom Uruguay

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.