FDA recall record

Wipro GE Healthcare Private GE Giraffe Warmer, Model/Catalog Number M1118179-01090979 recall — FDA Z-3080-2024

GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers.…

Class IIclassification
Z-3080-2024official record ID
Oct 2, 2024report date

Recall record at a glance

Official recordZ-3080-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmWipro GE Healthcare Private Ltd.
Report dateOct 2, 2024
Recall initiation dateAug 9, 2024
Quantity1
LocationBengaluru, India

Official product description

GE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.

Why the product was recalled

GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.

Distribution information

US and Australia, Denmark, France, Germany, Ireland, ITALY, New Zealand, Poland, Portugal, Slovenia, Spain, SWITZERLAND, United Arab Emirates, United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.