FDA recall record

Maquet Cardiovascular Fusion Bioline Vascular Graft, Part number M00201503046B0 recall — FDA Z-1245-2023

One (1) Fusion Bioline Vascular graft from lot 25162546, which failed guideline straightness testing, was intermingled with lot 25162549 and distributed…

Class IIclassification
Z-1245-2023official record ID
Mar 22, 2023report date

Recall record at a glance

Official recordZ-1245-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMaquet Cardiovascular, LLC
Report dateMar 22, 2023
Recall initiation dateJan 30, 2023
Quantity1 (US)
LocationWayne, NJ, United States

Official product description

Fusion Bioline Vascular Graft, Part number M00201503046B0

Why the product was recalled

One (1) Fusion Bioline Vascular graft from lot 25162546, which failed guideline straightness testing, was intermingled with lot 25162549 and distributed to customers. Use of the affected graft may result in reduction of blood flow or unwanted tension causing bleeding and the formation of a pseudo-aneurysm.

Distribution information

US Nationwide distribution in the states of FL, IN, LA, MI, MO, NC, NV, NY, PA, TN, TX, VA, WA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.