FDA recall record

Fujifilm Medical Systems U.S.A FUJINON ED-530XT. This device is intended for the recall — FDA Z-3225-2017

An update to the design and labeling was implemented to help reduce patient risk associated with inadequate reprocessing of the device. The action…

Class IIclassification
Z-3225-2017official record ID
Oct 4, 2017report date

Recall record at a glance

Official recordZ-3225-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmFujifilm Medical Systems U.S.A., Inc.
Report dateOct 4, 2017
Recall initiation dateJul 21, 2017
Quantity362 units
LocationWayne, NJ, United States

Official product description

FUJINON ED-530XT. This device is intended for the visualization of the duodenum and upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

Why the product was recalled

An update to the design and labeling was implemented to help reduce patient risk associated with inadequate reprocessing of the device. The action includes replacement of the forceps elevator mechanism, the O-ring seal, the distal end cap and issuance of a new Operation Manual.

Distribution information

Nationwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.