FDA recall record

FUJIFILM FDR Visionary Suite with CH-200 recall — FDA Z-0232-2018

A potential issue in the X-ray tube ceiling unit CH-200 which constitutes the Digital Radiography System FDR Visionary Suite. The X-ray tube assembly in…

Class IIclassification
Z-0232-2018official record ID
Dec 20, 2017report date

Recall record at a glance

Official recordZ-0232-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmFujifilm Medical Systems U.S.A., Inc.
Report dateDec 20, 2017
Recall initiation dateJun 15, 2017
Quantity3
LocationStamford, CT, United States

Official product description

FUJIFILM FDR Visionary Suite with CH-200. Model # CH-200, Catalog: Visionary Suite (FMSU Marketing Name) The FDR Visionary Suite is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations.

Why the product was recalled

A potential issue in the X-ray tube ceiling unit CH-200 which constitutes the Digital Radiography System FDR Visionary Suite. The X-ray tube assembly in these devices is mounted to a tube mounting flange, which is part of the tube holding shaft. where cracks may occur over time on the tube holding shaft near the base of the tube mounting flange.

Distribution information

US Distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records