FDA recall record

Encore Medical FOUNDATION KNEE SYSTEM NON STERILE Femoral A/P Cut Block - Type II REF recall — FDA Z-2262-2015

The markings on the distal face of the instruments are mis-oriented by 180. This includes Anterior and Posterior, which the surgeon uses to correctly…

Class IIclassification
Z-2262-2015official record ID
Aug 5, 2015report date

Recall record at a glance

Official recordZ-2262-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmEncore Medical, Lp
Report dateAug 5, 2015
Recall initiation dateJun 24, 2015
Quantity10 units
LocationAustin, TX, United States

Official product description

FOUNDATION KNEE SYSTEM NON STERILE Femoral A/P Cut Block - Type II REF 800-01-139

Why the product was recalled

The markings on the distal face of the instruments are mis-oriented by 180. This includes Anterior and Posterior, which the surgeon uses to correctly place the cut block on the resectioned face of the femur. The cut block is not symmetric, so correct orientation of the instrument in the A/P direction is critical to making proper anterior, posterior, and chamfer cuts to fit the femoral implant.

Distribution information

US Distribution to TX and GA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.