FDA recall record

Integra LifeSciences Footswitch accessory to the CUSA Clarity Ultrasonic Tissue Ablation recall — FDA Z-0694-2018

A faulty footswitch may result in unintended ultrasonic fragmentation activation. There are two scenarios when the failure mode may occur: 1) Ultrasonic…

Class IIclassification
Z-0694-2018official record ID
Mar 7, 2018report date

Recall record at a glance

Official recordZ-0694-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmIntegra LifeSciences Corp.
Report dateMar 7, 2018
Recall initiation dateSep 26, 2017
Quantity155
LocationPlainsboro, NJ, United States

Official product description

Footswitch accessory to the CUSA Clarity Ultrasonic Tissue Ablation System; Catalogue No: C7000 Product Usage: The footswitch is a component of the CUSA Clarity Ultrasonic Surgical Aspirator System, which is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft tissue is desirable. The CUSA Clarity Ultrasonic Surgical Aspirator is indicated for use in: Neurosurgery, Plastic and Reconstructive surgery, Orthopedic Surgery, Gynecological Surgery and Thoracic Surgery and the following specific uses are cleared for use in the US only: Gastrointestinal and Affiliated Organ Surgery including removal of benign or malignant tumors or other unwanted tissue, including hepatic parenchyma, in open or laparoscopic procedures, hepatic resection, tumor resection, lobectomy or trisegmentectomy, or removal of tissue during liver allotransplantation and donor hepatectomy Urological surgery including removal of renal parenchyma during nephrectomy or partial nephrectomy General Surgery including removal of benign or malignant tumors or other unwanted soft tissue in open or minimally invasive general surgical procedures Laparoscopic Surgery including removal of hepatic parenchyma in laparoscopic hepatic resection, lobectomy or trisegmentectomy, in laparoscopic donor hepatectomy or laparoscopic cholecystectomy or laparoscopic pancreatic jejunostomy, or pancreatectomy, or laparoscopic appendectomy, laparoscopic colon resection or laparoscopic partial gastrectomy.

Why the product was recalled

A faulty footswitch may result in unintended ultrasonic fragmentation activation. There are two scenarios when the failure mode may occur: 1) Ultrasonic fragmentation may be inadvertently activated at start-up or at the initial use once the footswitch is connected to the console and after the user primes the system, performs a handpiece test, and enters the Main Screen. 2) Ultrasonic fragmentation may remain activated during surgical use once the foot pedal is disengaged slowly.

Distribution information

Worldwide Distribution - US Nationwide in 33 states: AL, AR, CA, CO, CT, FL, GA, IA, ID, IL, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NM, NY, OH, OK, OR, PA, TN, TX, VA, WA, WI in the countries of Greece, South Africa, France, Germany, Italy, United Kingdom, Sweden, Netherlands, and Dominican Republic.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.