FDA recall record

Maquet Cardiovascular Us Sales Flow-i Anesthesia System Product Usage: The indication for recall — FDA Z-0804-2016

Electrical issues in the control, monitoring or panel sub-systems, caused by the printed circuit boards (PCBs) that control the sub-systems. The problems…

Class Iclassification
Z-0804-2016official record ID
Mar 2, 2016report date

Recall record at a glance

Official recordZ-0804-2016
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmMaquet Cardiovascular Us Sales, LLC
Report dateMar 2, 2016
Recall initiation dateDec 10, 2015
Quantity7 units within US
LocationWayne, NJ, United States

Official product description

Flow-i Anesthesia System Product Usage: The indication for the Flow-i Anesthesia System is administering inhalation anesthesia while controlling the entire ventilation of patients with no ability to breath, as well as in supporting patients with a limited ability to breath. The system is intended for use on neonatal to adult patient populations. The system is intended for use in hospital environments, except MRI environment by healthcare professionals trained in inhalation anesthesia administration.

Why the product was recalled

Electrical issues in the control, monitoring or panel sub-systems, caused by the printed circuit boards (PCBs) that control the sub-systems. The problems manifested as multiple re-starts of the sub-systems and, in a few instances, disabling of the system has occurred. Replacement PCBs are now available to correct the issues. Only the Model C20 is affected in the US.

Distribution information

US Nationwide in the states of CA, NY, and OR

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.