FDA recall record

Inpeco S.A FlexLab (FLX); Version: FLX-217-10 recall — FDA Z-0595-2026

The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous…

Class IIclassification
Z-0595-2026official record ID
Dec 3, 2025report date

Recall record at a glance

Official recordZ-0595-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmInpeco S.A.
Report dateDec 3, 2025
Recall initiation dateOct 30, 2025
Quantity1 unit (OUS only)
LocationLugano, N/A, Switzerland

Official product description

FlexLab (FLX); Version: FLX-217-10;

Why the product was recalled

The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results. This error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes.

Distribution information

US Domestic: MI. OUS International: Australia, Finland, France, Germany, Italy, Portugal, Spain, and UK.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceOct 15, 2025

Inpeco S.A FlexLab X (FXX) System recall — FDA Z-0107-2026

The FlexLab X System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for…

Class IIFDA: OngoingInpeco S.A.
Medical DeviceOct 15, 2025

Inpeco S.A FlexLab (FLX) System recall — FDA Z-0106-2026

The FlexLab System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for…

Class IIFDA: OngoingInpeco S.A.