Recall record at a glance
| Official record | Z-0353-2024 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | BioFire Diagnostics, LLC |
| Report date | Nov 29, 2023 |
| Recall initiation date | Oct 17, 2023 |
| Quantity | 5,406 systems |
| Location | Salt Lake City, UT, United States |
Official product description
FilmArray TORCH REF HTFA-ASY-0104. HTFA-ASY-0104 is the BIOFIRE TORCH Base which includes barcode scanner, touch screen computer and USB ports. The BioFire FilmArray Torch (BioFire Torch) is an automated in vitro diagnostic (IVD) device intended for use with FDA-cleared or approved IVD BioFire FilmArray Panels. Additional Part/Reference Numbers as part of the system: - HTFA-ASY-0102: BIOFIRE TORCH Duplex. This is a plastic component that covers the BIOFIRE TORCH Modules. The BIOFIRE TORCH Duplex is only necessary if the system includes more than two modules. - HFTA-SUB-0103: BIOFIRE TORCH Module. The BIOFIRE reagent pouches are inserted into the modules.
Why the product was recalled
Due to an increased risk of degradation of the power entry switch resulting from arcing inside of the power switch that may result in carbon build-up on the switch contacts. The carbon build-up may lead to excess heating inside of the power switch, subsequently leading to deformation of the power switch case. The deformation can create an opportunity for a power switch electrical short. In addition to an electrical short, deformation may result in an open circuit causing the power switch to fail. This event would only occur after the product is in use.
Distribution information
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, N, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, Virgin Island, VT, WA, WI, WV, and WY. The countries of Algeria, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Egypt, El Salvador, France, Germany, Guatemala, India, Indonesia, Italy, Japan, Korea, Kuwait, Malaysia, Mexico, New Zealand, Paraguay, Philippines, Poland, Qatar, Romania, Saudi Arabia, Singapore, Spain, Taiwan, Turkey, United Arab Emirates.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.