Recall record at a glance
| Official record | Z-2296-2020 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | BioFire Diagnostics, LLC |
| Report date | Jun 17, 2020 |
| Recall initiation date | Jan 22, 2020 |
| Quantity | 5,020 kits |
| Location | Salt Lake City, UT, United States |
Official product description
FilmArray BCID Panel, RFIT-ASY-0127, 6, Rx Only, CE, IVD, UDI: 00815381020093 used with the following BACT/ALERT Blood Culture Bottle Catalog No./Description: 410851/BACT/ALERT FA Plus 410852/BACT/ALERT FN Plus 410853/BACT/ALERT PF Plus 259790/BACT/ALERT SN Product Usage: is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with FilmArray systems.
Why the product was recalled
There is an increased risk of false positive Proteus results when using the Film Array Blood Culture Identification Panel (BCID) in vitro diagnostic test of blood cultures used in conjunction with specific media types. False positive Proteus results from this test may lead to patients receiving improper antibiotic treatment until culture results are available (within 24-48 hours).
Distribution information
Worldwide distribution - US Nationwide distribution and the countries of Argentina, Austria, Australia, Azerbaijan, Belgium, Bosnia-Herz., Botswana, Brazil, Bulgaria, Burkina Faso, Canada, Chile, Columbia, Costa Rica, Croatia, Czech Republic, Dominican Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Estonia, France, Gabon, Georgia, Germany, Greece, Guam, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Ivory Coast, Japan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Mexico, Morocco, Myanmar, Netherlands, Nicaragua, Oman, Pakistan, Panama, Paraguay, Peru, Phillipines, Poland, Portugal, Romania, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Sitzerland, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, United Arab Emirates, Vietnam.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.