FDA recall record

Stryker GmbH Femoral Nail, LEFT T2 GTN 8x300 mm recall — FDA Z-1939-2025

Investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but contained a right nail and another lot was…

Class IIclassification
Z-1939-2025official record ID
Jun 18, 2025report date

Recall record at a glance

Official recordZ-1939-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmStryker GmbH
Report dateJun 18, 2025
Recall initiation dateMay 22, 2025
Quantity1
LocationSelzach, Switzerland

Official product description

Femoral Nail, LEFT T2 GTN 8x300 mm. Catalog Number: 1850-0830S.

Why the product was recalled

Investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but contained a right nail and another lot was labeled as right version but contained a left nail.

Distribution information

Worldwide - US Nationwide distribution in the states of California and Missouri. The countries of Belgium, Japan, Netherlands.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.