Recall record at a glance
| Official record | Z-2757-2026 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Boston Scientific Corporation |
| Report date | Jul 29, 2026 |
| Recall initiation date | Jun 11, 2026 |
| Quantity | 193,000 units |
| Location | Saint Paul, MN, United States |
Official product description
FARADRIVE STEERABLE SHEATH CLEAR OUS, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PFCE21M402, IFU Reference Number 51659727-01
Why the product was recalled
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
Distribution information
Global Distribution. US Nationwide.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.