FDA recall record

Olympus Corporation of the Americas EZDilate Wire Guided Balloon- Indicated for use in adult recall — FDA Z-0225-2025

Mislabeled Glo Cath Label (attached to the device) states: 11 mm, 12 mm, 13 mm however the balloon is 8.5mm, 9.5mm and 10.5 mm and may result in prolonged…

Class IIclassification
Z-0225-2025official record ID
Oct 30, 2024report date

Recall record at a glance

Official recordZ-0225-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmOlympus Corporation of the Americas
Report dateOct 30, 2024
Recall initiation dateSep 13, 2024
Quantity209 units
LocationCenter Valley, PA, United States

Official product description

EZDilate Wire Guided Balloon- Indicated for use in adult and adolescent (> 12 years) populations to endoscopically dilate strictures of the alimentary tract. It is also indicated in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy. Model/Catalog Number: BD-410X-1055

Why the product was recalled

Mislabeled Glo Cath Label (attached to the device) states: 11 mm, 12 mm, 13 mm however the balloon is 8.5mm, 9.5mm and 10.5 mm and may result in prolonged surgery. All other labeling is correct for the balloon, including the Shelf Box with Front Box Label and Circular Star Label and the balloon Pouch

Distribution information

Worldwide Distribution: US distribution to states of: CA, FL, IL, NJ, NY, PA, TN, VA, WI; and Foreign OUS to countries of: CA, DE, IN, SG, Latin America.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.