FDA recall record

Qiagen Sciences EZ2 Connect MDx-IVD Designed to perform automated isolation and purification recall — FDA Z-1508-2024

Insufficient welding leads to the heater cable becoming brittle and present with an error: ERROR 3481 HEATER UNABLE TO REACH TEMPERATURE; the instrument…

Class IIclassification
Z-1508-2024official record ID
Apr 17, 2024report date

Recall record at a glance

Official recordZ-1508-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmQiagen Sciences LLC
Report dateApr 17, 2024
Recall initiation dateMar 4, 2024
Quantity168 units: 38 units US; 130 units OUS
LocationGermantown, MD, United States

Official product description

EZ2 Connect MDx-IVD Designed to perform automated isolation and purification of nucleic acids in molecular diagnostic and/or molecular biology applications. Model Number: 9003230

Why the product was recalled

Insufficient welding leads to the heater cable becoming brittle and present with an error: ERROR 3481 HEATER UNABLE TO REACH TEMPERATURE; the instrument may stop, and the sample being processed will be lost

Distribution information

Nationwide Foreign: AT, AU, BE, CH, CR, CZ, DE, DK, ES, FR, GB, HU, IT, JO, NL, NO, QA, SA, SE, SI, TH

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records