FDA recall record

Baxter Healthcare EXACTAMIX 1200 Compounder, model numbers 1200-DY, 1200-DX, 1200-DYR and recall — FDA Z-1963-2014

If the universal ingredient (UI) in an active configuration is changed using the Configuration Editor, a flush of the outlet pump tube will not be…

Class IIclassification
Z-1963-2014official record ID
Jul 9, 2014report date

Recall record at a glance

Official recordZ-1963-2014
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBaxter Healthcare Corp.
Report dateJul 9, 2014
Recall initiation dateJun 10, 2014
Quantity1758
LocationDeerfield, IL, United States

Official product description

EXACTAMIX 1200 Compounder, model numbers 1200-DY, 1200-DX, 1200-DYR and 1200-DXR; EM1200 DY Display, Replacement EM 1200 DY Display. EXACTAMIX 2400 Compounder model numbers 2400-DY, 2400-DX, 2400-DYR and 2400-DYX. Baxa Corporation. An automated pumping system that compounds multiple sterile ingredients into a finished solution in a single patient IV administration bag.

Why the product was recalled

If the universal ingredient (UI) in an active configuration is changed using the Configuration Editor, a flush of the outlet pump tube will not be initiated by the software. It could result in the original UI remaining in the tube which could be delivered into the next bag.

Distribution information

US Nationwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.