Draeger Medical Evita Infinity V500 Ventilator recall — FDA Z-0827-2016

The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly…

Class Iclassification
Z-0827-2016official record ID
Mar 2, 2016report date

Recall record at a glance

Official recordZ-0827-2016
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmDraeger Medical, Inc.
Report dateMar 2, 2016
Recall initiation dateFeb 4, 2016
Quantity2553
LocationTelford, PA, United States

Official product description

Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.

Why the product was recalled

The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.

Distribution information

US Nationwide Distribution including Puerto Rico.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceDec 25, 2024

Atlan A300XL recall — FDA Z-0668-2025

The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

Class IFDA: OngoingDraeger Medical, Inc.
Medical DeviceDec 25, 2024

Atlan A300 recall — FDA Z-0667-2025

The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

Class IFDA: OngoingDraeger Medical, Inc.
Medical DeviceNov 20, 2024

Atlan A350XL recall — FDA Z-0305-2025

The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

Class IFDA: OngoingDraeger Medical, Inc.
Medical DeviceNov 20, 2024

Atlan A350 recall — FDA Z-0304-2025

The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

Class IFDA: OngoingDraeger Medical, Inc.