FDA recall record

Olympus Corporation of the Americas EVIS EXERA Duodenovideoscope Olympus TJF Type 160F/VF recall — FDA Z-1845-2016

New reprocessing instructions for the Olympus TJF-160/VF duodenoscopes, consisting of revised manual cleaning and high level disinfection procedures they…

Class IIclassification
Z-1845-2016official record ID
Jun 8, 2016report date

Recall record at a glance

Official recordZ-1845-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmOlympus Corporation of the Americas
Report dateJun 8, 2016
Recall initiation dateMar 15, 2016
QuantityN/A
LocationCenter Valley, PA, United States

Official product description

EVIS EXERA Duodenovideoscope Olympus TJF Type 160F/VF used with an Olympus video system center, light source, documentation equipment monitors, Endo Therapy accessories and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.

Why the product was recalled

New reprocessing instructions for the Olympus TJF-160/VF duodenoscopes, consisting of revised manual cleaning and high level disinfection procedures they plan to implement as well as a new cleaning brush (MAJ-1534), which is enclosed with the notification.

Distribution information

Nationwide:

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.