Recall record at a glance
| Official record | Z-2048-2016 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Ev3, Inc. |
| Report date | Jun 29, 2016 |
| Recall initiation date | May 23, 2016 |
| Quantity | 2 |
| Location | Plymouth, MN, United States |
Official product description
EV3 EverFiex" Self-expanding Peripheral Stent System, Sterilized using Ethylene Oxide, Rx Only, PRB35-08-040-080. The Protege EverFiex" Self-expanding Peripheral Stent System is a self-expanding Nitinol stent system intended for permanent implantation. The self-expanding stent is made of a nickel titanium alloy (Nitinol) and comes pre-mounted on a 6 F, 0.035" over-the-wire delivery system. The stent is cut from a Nitinol tube in an open lattice design, and has tantalum radiopaque markers at the proximal and distal ends of the stent. Upon deployment, the stent achieves its predetermined diameter and exerts a constant, gentle outward force to establish patency
Why the product was recalled
Medtronic is issuing a medical device recall to return product potentially associated with an outer box mislabel issue for two Protege EverFiex" Self-expanding Peripheral Stent Systems, model PRB35-08-040- 080, from lot A164580. The size on the outer box is labeled as 8x40x80mm while potentially containing a 6x40x120mm device as indicated by both the inner pouch and device labels. This potential issue is limited to two EverFiex stent system devices from lot A164580 that were repackaged by Medtronic and distributed. This does not affect any other Medtronic product or implantable devices.
Distribution information
Nationwide Distribution to LA and WI.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.