FDA recall record

LivaNova Deutschland GmbH Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software recall — FDA Z-0571-2025

Heart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through…

Class IIclassification
Z-0571-2025official record ID
Dec 4, 2024report date

Recall record at a glance

Official recordZ-0571-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmLivaNova Deutschland GmbH
Report dateDec 4, 2024
Recall initiation dateOct 18, 2024
Quantity38
LocationMunich, Germany

Official product description

Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software

Why the product was recalled

Heart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through backup control panel, but second reset may occur, may lead to gas blender standby, then gas blender must be reactivated using user interface on blender, sensors deactivated until new case started; GUI reset can lead to Hypoxia and Hypoperfusion.

Distribution information

US Nationwide distribution in the states of CA, VA, GA, IN, NJ, FL, CO, MS, NV.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.