FDA recall record

Respironics California Esprit V1000 Ventilator and Esprit/V200 Conversion Ventilator The recall — FDA Z-0485-2015

If the power supply fan mounting screws are installed incorrectly, the ends of the screws may touch the power supply printed circuit board assembly (PCBA)…

Class IIclassification
Z-0485-2015official record ID
Dec 3, 2014report date

Recall record at a glance

Official recordZ-0485-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRespironics California Inc.
Report dateDec 3, 2014
Recall initiation dateOct 29, 2014
Quantity3,943 units total (3,071 units in US)
LocationCarlsbad, CA, United States

Official product description

Esprit V1000 Ventilator and Esprit/V200 Conversion Ventilator The Esprit V1000 and Esprit/V200 Conversion Ventilators are microprocessor-controlled, electrically powered mechanical ventilators. Intended for use by medical personnel to provide continuous or intermittent ventilator support for adult, pediatric and neonatal patients.

Why the product was recalled

If the power supply fan mounting screws are installed incorrectly, the ends of the screws may touch the power supply printed circuit board assembly (PCBA) and short to ground. This may prevent the ventilator to switch to battery power if AC power is disrupted.

Distribution information

Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Ecuador, Egypt, Greece, Hong Kong, India, Italy, Korea, Lebanon, Malaysia, Mexico, Pakistan, Panama, Peru, Philippines, Saudi Arabia, Singapore, South Africa, Spain, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Vietnam, and Yemen.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.