Recall record at a glance
| Official record | Z-0123-2022 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Philips Ultrasound Inc. |
| Report date | Oct 27, 2021 |
| Recall initiation date | Sep 21, 2021 |
| Quantity | 107 |
| Location | Bothell, WA, United States |
Official product description
EPIQ CVxi Diagnostic Ultrasound Systems, Product Code: 795232
Why the product was recalled
Ultrasound system software issue can cause an EchoNavigator error notification, leading to an unresponsive system, necessitating system restart, which could lead to therapy/treatment delay and/or unnecessary therapy/treatment. Can occur during ultrasound-guided TEE with EchoNavigator, if secondary screen capture is enabled and user presses other system buttons before secondary capture completed.
Distribution information
US: MI, AZ, IA, TN, OH, CO, WI, NY, NJ, NC, LA, IL, CA, KY, IN, MD, TX, PA. OUS: Denmark, Latvia, Belgium, Hungary, Argentina, Canada, Spain, Germany, Greece, Korea, Republic of Japan, France, Australia, Egypt, Switzerland, Italy, Senegal, Israel, Indonesia, Netherlands, Hong Kong, United Kingdom, Poland, Finland, India, New Zealand
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.