EOS Imaging EOS System X- ray beam Digital radiography system used in general radiographic recall — FDA Z-0529-2014

EOS imaging discovered during production internal testing that the X-ray beam collimation set up by the operator at the user interface of the acquisition…

Class IIclassification
Z-0529-2014official record ID
Jan 15, 2014report date

Recall record at a glance

Official recordZ-0529-2014
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmEOS Imaging
Report dateJan 15, 2014
Recall initiation dateNov 7, 2013
Quantity22 EOS systems are installed in US
LocationParis France, France

Official product description

EOS System X- ray beam Digital radiography system used in general radiographic examinations.

Why the product was recalled

EOS imaging discovered during production internal testing that the X-ray beam collimation set up by the operator at the user interface of the acquisition workstation is different from the actual collimation of the X-ray beam operated by the EOS system during the X-ray acquisition.

Distribution information

Nationwide Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceSep 26, 2018

EOS Imaging sterEOS workstation recall — FDA Z-3188-2018

3D projections may be incorrect when the pair of images used is a secondary capture generated from a series containing more than two images, where one of the images has been manually…

Class IIFDA: TerminatedEOS Imaging