Recall record at a glance
| Official record | Z-0806-2025 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Elucent Medical Inc. |
| Report date | Jan 15, 2025 |
| Recall initiation date | Nov 18, 2024 |
| Quantity | 283 (48 units) |
| Location | Eden Prairie, MN, United States |
Official product description
EnVisio Patient Pad Transition Cushion, Model/Catalog Number: ENVISIO-SYS Product Description: The EnVisio Patient Pad Transition Cushion is a component of the EnVisio Navigation System which is a medical device that detects the presence of the SmartClip Soft Tissue Marker(s) and provides real-time three-dimensional navigation during surgery.
Why the product was recalled
There is a potential safety risk caused by physical disassociation of the back section from the bed frame of Skytron model series 3600, 6500, 6600 and 6700 operating room beds when used with the EnVisio Navigation System.
Distribution information
US Distribution: AZ, CA, CO, CT, FL, GA, IL, KY, MN, NC, ND, NJ, NY, OH, PA, SC, WI
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.