FDA recall record

Xeridiem Mediem Medical Devices EndoVive 3s Low Profile Balloon Kits Part Number: M00549550 recall — FDA Z-0987-2016

A feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is…

Class IIclassification
Z-0987-2016official record ID
Mar 2, 2016report date

Recall record at a glance

Official recordZ-0987-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmXeridiem Mediem Medical Devices Inc.
Report dateMar 2, 2016
Recall initiation dateDec 23, 2015
Quantity38
LocationTucson, AZ, United States

Official product description

EndoVive 3s Low Profile Balloon Kits Part Number: M00549550 (XMD P/N 70-0050-B24) The EndoVive 3s Low Profile Balloon is intended to provide gastric access through a gastrointestinal stoma tract for enteral feeding, medication administration and gastric decompression. The EndoVive 3s Low Profile Balloon is indicated for use in adult and pediatric patients who require long-term feeding, are unable to tolerate oral feeding, are at low risk for aspiration or require gastric decompression and/or medication delivery directly into the stomach.

Why the product was recalled

A feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric contents to leak out of the device in some instances.

Distribution information

Nationwide distribution to MA only.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records