Recall record at a glance
| Official record | Z-0201-2023 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Covidien, LP |
| Report date | Nov 23, 2022 |
| Recall initiation date | Sep 28, 2022 |
| Quantity | 30658 |
| Location | North Haven, CT, United States |
Official product description
Endo Stitch V-Loc PBT Non-Absorbable Reload, Models VLOCN004L, VLOCN006L, VLOCN008L, VLOCN204L, VLOCN206L, VLOCN208L, VLOCN304L, VLOCN306L, VLOCN308L
Why the product was recalled
Medtronic has received 210 reports of needles breaking during endoscopic suturing. The potential harm(s) include a delay in treatment/therapy, the potential for a portion of the needle to remain in the patient, foreign body reaction, allergic reaction, tissue injury and unintended radiation exposure from additional imaging for both intra-op and postoperative scenarios.
Distribution information
Domestic distribution nationwide. Foreign distribution to Australia Austria Belgium Canada Canary Islands Chile Croatia Cyprus Denmark Egypt Finland France Germany Greece Hungary Ireland Israel Italy Jordan Kazakhstan Kuwait Lithuania Luxembourg Mexico Netherlands New Zealand Norway Poland Portugal Puerto Rico Qatar Reunion Romania Russian Federation Saudi Arabia Slovakia South Africa Spain Sweden Switzerland Turkey United Arab Emirates United Kingdom
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.