FDA recall record

GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING Elscint Varicam; dual-head nuclear medicine recall — FDA Z-0294-2026

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated…

Class IIclassification
Z-0294-2026official record ID
Oct 29, 2025report date

Recall record at a glance

Official recordZ-0294-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmGE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Report dateOct 29, 2025
Recall initiation dateSep 12, 2025
Quantity15 units
LocationHaifa, Israel

Official product description

Elscint Varicam; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Why the product was recalled

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Distribution information

Worldwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.