FDA recall record

Laserex Systems Ellex Tango Reflex with slit lamp Laser Ophthalmic In the SLT mode, the recall — FDA Z-2172-2017

It was discovered the unit produced a laser emission without pressing the fire button when the slit-lamp was driven to its lowest position prior to use on…

Class IIclassification
Z-2172-2017official record ID
Jun 14, 2017report date

Recall record at a glance

Official recordZ-2172-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmLaserex Systems Inc.
Report dateJun 14, 2017
Recall initiation dateApr 4, 2017
Quantity83
LocationEDEN PRAIRIE, MN, United States

Official product description

Ellex Tango Reflex with slit lamp Laser Ophthalmic In the SLT mode, the device is intended to be used for selective laser trabeculoplasty (ST) operations (laser trabeculoplasty for primary open angle glaucoma). In the VAG mode it is intended to be used to perform procedures requiring the rupture of tissue in the eye for Iridotomy and Iridectomy, Posterior capsulotomy and Posterior membranectomy.

Why the product was recalled

It was discovered the unit produced a laser emission without pressing the fire button when the slit-lamp was driven to its lowest position prior to use on any patient.

Distribution information

US Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: FDA Mandated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.