FDA recall record

Cytrellis Biosystems ellacor System with Micro-Coring Technology-Indicated for use by recall — FDA Z-2393-2023

Potential failure of a bearing adhesive joint that can occur due to an assembly issue of the X/Y stage of the handpiece. If a broken glue bond were to…

Class IIclassification
Z-2393-2023official record ID
Aug 16, 2023report date

Recall record at a glance

Official recordZ-2393-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmCytrellis Biosystems, Inc.
Report dateAug 16, 2023
Recall initiation dateJun 23, 2023
Quantity127 systems
LocationWoburn, MA, United States

Official product description

ellacor System with Micro-Coring Technology-Indicated for use by medical professionals for the treatment of moderate and severe wrinkles in the mid and lower face in adults aged 22 years or older with Fitzpatrick skin types I-IV. Model #: Gen 4

Why the product was recalled

Potential failure of a bearing adhesive joint that can occur due to an assembly issue of the X/Y stage of the handpiece. If a broken glue bond were to occur, it could cause an overlap of cores, potentially leading to prolonged healing and irregularities in patient skin texture post-treatment, and/or scarring

Distribution information

Nationwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.