FDA recall record

Mortara Instrument ELI 380 Electrocardiograph recall — FDA Z-1796-2015

Mortara Instrument, Inc. has recently become aware of a potential safety hazard involving our ELI 380 electrocardiograph. When used in a particular…

Class IIclassification
Z-1796-2015official record ID
Jun 24, 2015report date

Recall record at a glance

Official recordZ-1796-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMortara Instrument, Inc.
Report dateJun 24, 2015
Recall initiation dateMay 29, 2015
Quantity176
LocationMilwaukee, WI, United States

Official product description

ELI 380 Electrocardiograph. The ELI 380 is intended to be a high-performance, multichannel resting electrocardiograph. As a resting electrocardiograph, the ELI 380 simultaneously acquires data from each lead. Once the data is acquired, it can be analyzed, reviewed, stored, printed or transmitted. It is a device primarily intended for use in hospitals, but may be used in medical clinics and offices of any size.

Why the product was recalled

Mortara Instrument, Inc. has recently become aware of a potential safety hazard involving our ELI 380 electrocardiograph. When used in a particular workflow, acquired ECG waveforms for one patient may become associated with the patient demographics for a different patient when the record is transmitted to a records management system.

Distribution information

Worldwide Distribution - US Distribution to the states of : AZ, CA, FL, GA, IL, MI, NE, SC, UT, WA and WI., and to the countries of : CHINA, PORTUGAL, CANADA, OMAN, MALAYSIA, NEW ZEALAND, INDIA, POLAND, LEBANON, SWITZERLAND, FINLAND, ROMANIA,UNITED ARAB EMIRATES, UNITED KINGDOM, AUSTRALIA, THE NETHERLANDS, ITALY, FRANCE, GERMANY, THAILAND,SAUDI, ARABIA, NORWAY and IRELAND.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.