FDA recall record

Elekta Monaco - Product Usage: used to make treatment plans for patients with prescriptions recall — FDA Z-2563-2021

Contour changes can be saved on an unintended image set. In addition, these contour edits do not cause the dose to be frozen on the plans associated with…

Class IIclassification
Z-2563-2021official record ID
Oct 6, 2021report date

Recall record at a glance

Official recordZ-2563-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmElekta Inc.
Report dateOct 6, 2021
Recall initiation dateSep 14, 2021
Quantity214 units
LocationSaint Charles, MO, United States

Official product description

Elekta Monaco - Product Usage: used to make treatment plans for patients with prescriptions for external beam radiation therapy.

Why the product was recalled

Contour changes can be saved on an unintended image set. In addition, these contour edits do not cause the dose to be frozen on the plans associated with that image set.

Distribution information

Worldwide distribution - US Nationwide distribution in the states of PA, MI, WI, NY, TN, NJ, IA, TX and the countries of United Kingdom, Turkey, Switzerland, Sweden, Spain, Netherlands, Korea, Thailand, Japan, Italy, Hong Kong, Germany, France, Denmark, China, Canada, Belgium, Bahrain, Australia.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.