Boston Scientific EKOS Control System (CU) 4.0 recall — FDA Z-0637-2021

This is a follow-up to the EKOS Field Safety Notice dated April 30, 2019. The Radio Frequency (RF) board has been redesigned to prevent occasional…

Class IIclassification
Z-0637-2021official record ID
Dec 23, 2020report date

Recall record at a glance

Official recordZ-0637-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmBoston Scientific Corporation
Report dateDec 23, 2020
Recall initiation dateNov 3, 2020
Quantity90 (OUS)
LocationMarlborough, MA, United States

Official product description

EKOS Control System (CU) 4.0. UPN 600-40500 The EkoSonic MACH4 Endovascular (EKOS) System consists of a single use Infusion Catheter and Ultrasonic Core, and a reusable Control Unit 4.0. The device delivers the physician-specified fluids and ultrasound to the intravascular treatment site. The reusable Control Unit provides power to the device and provides the user interface for operator control. A reusable, non-sterile connector interface cable (CIC) connects the Control Unit to the Ultrasonic Core and Infusion Catheter.

Why the product was recalled

This is a follow-up to the EKOS Field Safety Notice dated April 30, 2019. The Radio Frequency (RF) board has been redesigned to prevent occasional internal board failures at lower temperatures. All EKOS CU 4.0 consoles containing the old RF board design will be replaced with the newly designed CU 4.0 consoles.

Distribution information

No US distribution. Foreign distribution to Latin America and EU.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.