FDA recall record

COVIDIEN EEA Circular Stapler with Tri-Staple Technology, Part No recall — FDA Z-0070-2019

The firm identified the potential for a device to have an incorrect tissue gap. Use of a device with an incorrect tissue gap may result in incomplete…

Class IIclassification
Z-0070-2019official record ID
Oct 10, 2018report date

Recall record at a glance

Official recordZ-0070-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCOVIDIEN LLC
Report dateOct 10, 2018
Recall initiation dateAug 17, 2018
Quantity1,902
LocationMansfield, MA, United States

Official product description

EEA Circular Stapler with Tri-Staple Technology, Part No. TRIEEA28MT TRIEEA28XT TRIEEA31MT TRIEEA31XT Product Usage: The EEA Circular Stapler with Tri-Staple Technology places a circular, triple staggered row of titanium staples and resects the excess tissue, creating a circular anastomosis.

Why the product was recalled

The firm identified the potential for a device to have an incorrect tissue gap. Use of a device with an incorrect tissue gap may result in incomplete staple formation and/or the inability to remove the device from tissue following application potentially leading to bleeding, anastomotic leak or tissue trauma. There have been no reports of serious injury related to this issue.

Distribution information

Worldwide Distribution - US Nationwide in the states of CA, FL, IL, NC, NY, UT. and the countries of Austria, Belgium, Denmark, Finland, France, Germany, Iceland, Italy, Latvia, Netherlands, Norway, Poland, Portugal, Reunion, Spain, Sweden, Switzerland, and United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.