FDA recall record

Ameditech EDI BZO Uncut Sheet - Cup*, Item No. 160170 recall — FDA Z-0501-2016

Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which…

Class IIIclassification
Z-0501-2016official record ID
Dec 30, 2015report date

Recall record at a glance

Official recordZ-0501-2016
CategoryMedical Device Recalls
ClassificationClass III
FDA record statusTerminated
Recalling firmAmeditech Inc.
Report dateDec 30, 2015
Recall initiation dateNov 16, 2015
Quantity45,584 kits
LocationSan Diego, CA, United States

Official product description

EDI BZO Uncut Sheet - Cup*, Item No. 160170

Why the product was recalled

Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.

Distribution information

Nationwide Distribution.

How to read the classification

FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceOct 7, 2020

Ameditech Noble 1 Step+ Cup recall — FDA Z-3039-2020

Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

Class IIFDA: TerminatedAmeditech Inc.
Medical DeviceOct 7, 2020

Ameditech TDDA Drugs of Abuse Cup recall — FDA Z-3044-2020

Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

Class IIFDA: TerminatedAmeditech Inc.