FDA recall record

Varian Medical Systems Eclipse(TM) Treatment Planning System with Proton Convolution recall — FDA Z-0771-2019

There is an anomaly with the Eclipse(TM) Treatment Planning System [TPS] Proton Convolution Superposition [PCS] dose calculation algorithm. The PCS algo-…

Class IIclassification
Z-0771-2019official record ID
Jan 30, 2019report date

Recall record at a glance

Official recordZ-0771-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmVarian Medical Systems, Inc.
Report dateJan 30, 2019
Recall initiation dateDec 26, 2018
Quantity1450 devices
LocationPalo Alto, CA, United States

Official product description

Eclipse(TM) Treatment Planning System with Proton Convolution Superposition algorithm license only The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, elec- tron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.

Why the product was recalled

There is an anomaly with the Eclipse(TM) Treatment Planning System [TPS] Proton Convolution Superposition [PCS] dose calculation algorithm. The PCS algo- rithm calculates the water equivalent range incorrectly for non-square 3D CT images (either different number of pixels in X and Y, or non-square pixels). Before the dose is calculated, Eclipse resamples the CT images to create a calculation image with a maximum resolution of 256 x 256. The PCS algorithm assumes uniform resolution in X and Y directions, or that X = Y for all images, and erroneously sets the Y = X for dose calcu- lation. For images that are not square, where either X<>Y, or the length of X does not equal the length of Y, the computed water equivalent range R is erroneous compared to the correct range R: 1) X=Y; R =R (R is correct); 2) X<Y: R <R (R is too small) and 3) X>Y R >R (R is too big).

Distribution information

Worldwide distribution: US (nationwide) and countries of: Australia, Brazil, Denmark France, Germany, Japan, Poland, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Sweden, Switzerland, Taiwan, and United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceMay 13, 2020

Varian Multileaf Collimator [MLC] recall — FDA Z-1899-2020

After a recent upgrade to the collimator software version 8.5, the firm became aware that the multi-leaf collimator leaves did not move during an arc treatment. The issue is related to a…

Class IIFDA: TerminatedVarian Medical Systems, Inc.