FDA recall record

Philips Visicu eCareCoordinator allows the customer to schedule patient tasks recall — FDA Z-1557-2016

The adherence check generates a software error. Two issues occur as a result of the software error: 1) No adherence flag is generated for that day; 2) No…

Class IIclassification
Z-1557-2016official record ID
May 4, 2016report date

Recall record at a glance

Official recordZ-1557-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmPhilips Visicu
Report dateMay 4, 2016
Recall initiation dateMar 10, 2016
Quantity9
LocationBaltimore, MD, United States

Official product description

eCareCoordinator allows the customer to schedule patient tasks (i.e., take weight measurement, take blood pressure). For medical use by professional medical staff.

Why the product was recalled

The adherence check generates a software error. Two issues occur as a result of the software error: 1) No adherence flag is generated for that day; 2) No task is generated for the date 14 days later. An error message is generated and forwarded to operations. Subsequently, for days in which no task was generated, patient will not get a reminder on the tablet to take a measurement. If the pa

Distribution information

US Distribution -- to the stated of FL, MA, MI, MS, PA, and KS.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.