FDA recall record

NDDd Medical Technologies EasyOne Pro LAB VOS Respiratory Analysis System Pulmonary Function recall — FDA Z-0597-2020

The CO/CO2 sensor cable harness may have lower electromagnetic shielding, which might have resulted in a non-compliance with the IEC 60601-1-2 standard.…

Class IIclassification
Z-0597-2020official record ID
Dec 11, 2019report date

Recall record at a glance

Official recordZ-0597-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmNDDd Medical Technologies, Inc.
Report dateDec 11, 2019
Recall initiation dateAug 15, 2017
Quantity37 total devices
LocationAndover, MA, United States

Official product description

EasyOne Pro LAB VOS Respiratory Analysis System Pulmonary Function Testing Device, Model 3100-1

Why the product was recalled

The CO/CO2 sensor cable harness may have lower electromagnetic shielding, which might have resulted in a non-compliance with the IEC 60601-1-2 standard. As a consequence of a lower electromagnetic shielding provided by the cable harness, the CO/CO2 sensor could have been influenced by high frequencies emitted by very close devices.

Distribution information

The products were distributed to the following US states: KY, MO, NY, OH, PA

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.