FDA recall record

Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, running recall — FDA Z-1926-2025

A new motor/encoder and cable was introduced into instruments manufactured starting with serial number 300141 incorrectly. This incorrect installation is…

Class IIclassification
Z-1926-2025official record ID
Jun 18, 2025report date

Recall record at a glance

Official recordZ-1926-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmBeckman Coulter, Inc.
Report dateJun 18, 2025
Recall initiation dateApr 23, 2025
Quantity324 units
LocationChaska, MN, United States

Official product description

DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, running software versions 1.20

Why the product was recalled

A new motor/encoder and cable was introduced into instruments manufactured starting with serial number 300141 incorrectly. This incorrect installation is introducing error codes for the Dxl 9000 which prevents the assay from being performed and a delay in results from being generated, and may require field service to resolve the delay.

Distribution information

Worldwide distribution: US (Nationwide) and OUS Global (Foreign) to countries of: Italy, Germany, Spain, New Zealand, France, Australia, Slovakia, Switzerland, Ireland, Brazil, Taiwan, Province of China, India, United Kingdom of Great Britain and Northern Ireland, Hungary, Austria, Portugal, Croatia, Kazakhstan, Cote d'lvoire, Poland, Korea, Republic of, Czechia, Algeria, Libya, Oman, Singapore, and Belgium.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.