Recall record at a glance
| Official record | Z-1326-2025 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Beckman Coulter, Inc. |
| Report date | Mar 19, 2025 |
| Recall initiation date | Nov 13, 2024 |
| Quantity | 245 units |
| Location | Chaska, MN, United States |
Official product description
DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137
Why the product was recalled
The DxI 9000 Access Immunoassay Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer Error that is displayed on the navigation bar in the User Interface. The analyzer enters the red system status and will generate a SYS flag. The tests that are in progress may be cancelled and may cause a delay in reporting patient results
Distribution information
US Nationwide. Global Distribution.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.