Recall record at a glance
| Official record | Z-2602-2021 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Beckman Coulter Inc. |
| Report date | Oct 6, 2021 |
| Recall initiation date | Jul 21, 2021 |
| Quantity | 123 systems |
| Location | Brea, CA, United States |
Official product description
DxA 5000 (DxA Automation System, DxA 5000 fit), "For in Vitro Diagnostic Use", Reference Numbers: B87352 with B57634, UDI: 15099590707118
Why the product was recalled
There is a potential that sample carriers (w or w/o samples) within automated sampling system may transport samples to the outer lane which does not contain a physical barrier to restrain samples. Samples are dropped when they reach the end of the outer lane which could result in Delay in providing patient with appropriate medical treatment. New sample needs to be drawn. Substantial discharge of biohazards material - Infection with a human transmissible blood-borne or animal pathogen which could result in Infection with a human transmissible blood-borne pathogen (e.g. infectious HBV, HCV, HIV).
Distribution information
US Nationwide distribution in the states of AL, CA, IL, In, KY, MN, MI, OH, PA, TX, WV.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.