Agfa N.V DX-D 600 Product Usage: The DX-D 600 system is a ceiling mounted General recall — FDA Z-3245-2018

A supplier for the Agfa system noted a potential for the steel support cables of a similar device to break. This led to a mandatory review of all systems,…

Class IIclassification
Z-3245-2018official record ID
Oct 3, 2018report date

Recall record at a glance

Official recordZ-3245-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAgfa N.V.
Report dateOct 3, 2018
Recall initiation dateAug 15, 2018
Quantity108
LocationMortsel, Belgium

Official product description

DX-D 600 Product Usage: The DX-D 600 system is a ceiling mounted General Radiography X-Ray imaging system used in hospitals, clinics and medical practices by physicists, radiographers and radiologists to make, process and view static X-Ray radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts on adult, pediatric or neonatal patients.

Why the product was recalled

A supplier for the Agfa system noted a potential for the steel support cables of a similar device to break. This led to a mandatory review of all systems, changing the pulley and cables if any deterioration in the cables or irregular or excessive pulley wear. If the two cables that support the equipment to its roof anchor break, the equipment might fall and cause harm to the patient, or others.

Distribution information

Worldwide Distribution - US Nationwide in the states of CA, DC, FL, IL, IN, KS, LA, MA, MN, MO, NV, NY, OH, OR, SC and TN. Foreign distribution to Canada.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.