FDA recall record

Boston Scientific Neuromodulation Driver Instrument, part of the Superion Indirect recall — FDA Z-1972-2023

Pending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver…

Class IIclassification
Z-1972-2023official record ID
Jun 21, 2023report date

Recall record at a glance

Official recordZ-1972-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmBoston Scientific Neuromodulation Corporation
Report dateJun 21, 2023
Recall initiation dateMay 12, 2023
Quantity21,533
LocationValencia, CA, United States

Official product description

Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Superion Indirect Decompression System IFU, Superion IDS Kit IFU, Surgical Technique Manual

Why the product was recalled

Pending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver instrument tip breaks, which may result in metal fragments (Driver teeth/tips) within the implant location; and if metal fragments are not removed and remain in situ, MRI scans are NOT advised due to potential risk of patient injury.

Distribution information

US Nationwide Distribution: IN, NY, IL, PA, FL, SC, TX, MA, VA, AL, WI, OK, MO, MI, OH, WV, NC, UT, NH, CA, NJ, TN, NV, AZ, NE, IA, CT, ME, LA, KY, MS, MD, DC, GA, CO, KS, ID, MN, AR, HI, NM, WA, SD, OR, DE, WY, AK, and ND

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.