FDA recall record

Breg Dressing Rect L, P/N 02428 Product Usage: These products are sterile dressings used as recall — FDA Z-1903-2017

They may not have been sealed prior to sterilization. Products with unsealed pouches will be non-sterile.

Class IIclassification
Z-1903-2017official record ID
May 10, 2017report date

Recall record at a glance

Official recordZ-1903-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBreg Inc.
Report dateMay 10, 2017
Recall initiation dateMar 30, 2017
Quantity2,609 units total
LocationCarlsbad, CA, United States

Official product description

Dressing Rect L, P/N 02428 Product Usage: These products are sterile dressings used as an insulation barrier between the Polar Care cold therapy pad and the skin. Sterile dressings are typically applied in a non-sterile environment after post-op and on occasion may be applied in a sterile environment. Breg has determined that a small percentage (approximately 0.22%) of these products may have been sent with unsealed pouches by the contract manufacturer. Customers will be contacted and instructed to inspect their inventory for any product with unsealed pouches and discard them.

Why the product was recalled

They may not have been sealed prior to sterilization. Products with unsealed pouches will be non-sterile.

Distribution information

US Nationwide Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.