FDA recall record

Covidien Dover 100% Silicone Foley Catheter with Temperature Sensor 14FR 5CC2W, Product recall — FDA Z-1984-2020

Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient…

Class IIclassification
Z-1984-2020official record ID
May 20, 2020report date

Recall record at a glance

Official recordZ-1984-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCovidien LLC
Report dateMay 20, 2020
Recall initiation dateApr 30, 2020
Quantity468 units
LocationMansfield, MA, United States

Official product description

Dover 100% Silicone Foley Catheter with Temperature Sensor 14FR 5CC2W, Product Code: 50514T - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.

Why the product was recalled

Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment

Distribution information

US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records